Validation study demonstrates prospective detection of pressure-induced sensor drops in continuous glucose monitoring, suggesting potential to prevent false hypoglycemia alarms.
BACKGROUND: Continuous glucose monitoring (CGM) sensors are vulnerable to pressure-induced sensor attenuations (PISAs). Pressure-induced sensor attenuations are characterized by a rapid but erroneous drop in blood glucose (BG) readings, and if these BG readings are hypoglycemic or near-hypoglycemic they can cause unwarranted actions to be taken (eg, triggering of hypoglycemia alarms, shut-off of an insulin pump). In this work, we propose a prospective algorithm (PA) to detect the onset of PISAs where the minimum sensor-reported BG value during the PISA is less than 85 mg/dL. METHODS: The PA is composed of a screening module and a probability estimation module (PEM). The PEM was trained on a data set with 846 PISAs (59 122 hours of data, 67 patients). The optimal version of the PA was selected by assessing performance of each version of the PA on a data set with 89 PISAs (8189 hours of data, 11 patients). Given the highly imbalanced nature of the problem (baseline prevalence of PISAs onset in the testing set is 1.2%), performance was primarily assessed using precision-recall (PR) curves. RESULTS: The final version of the PA was tested on a data set with 325 PISAs (26 890 hours of data, 33 patients), resulting in an area under the PR curve of 0.590 (95% confidence interval [0.509, 0.672]), indicating excellent performance against the baseline occurrence. CONCLUSIONS: The results indicate that the PA could feasibly be used to inform other decision-making processes, for example, canceling of hypoglycemia alarms that would otherwise be triggered.
No takes yet. Share an insight, caveat, or question.
Benjamin Lobo (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: