We describe a simple near-infrared hygrometer based on a broadband light-emitting diode (LED) centered at 1320 nm. Light from the LED is collimated and passes through a rotating optical-filter assembly and a sample cell. The water vapor concentration is measured using a synchronous detection scheme. This compact and robust instrument has a response time of 0.15 s and can reliably detect short-term absolute humidity changes as small as 0.6 mg/L (which equate to 0.02% absorption); the long-term stability of the absolute humidity measurement over a period of 4 days is approximately 1 mg/L. These performance characteristics are particularly well suited to measurements associated with human respiration, and we present results in this humidity regime.
No takes yet. Share an insight, caveat, or question.
Wilson et al. (1995) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: