In situ and in vivo treatment plan verification and beam monitoring as well as dose control during heavy-ion tumour therapy can be performed in principle by measurements of range distributions of beta(+)-emitting nuclei by means of PET techniques. For this purpose the performance of different types of positron camera as well as the results of in-beam PET experiments using beams of beta(+)-active heavy ions (15O, 17F and 19Ne with energies of 300-500 A MeV) are presented. Following the deduced performance requirements a PET scanner that is designed for clinical use in experimental heavy-ion therapy at GSI Darmstadt has been built. This limited angle tomograph consists of two large-area detector heads based on position sensitive BGO detectors and is predicted to perform the measurement of the end point of a beta(+)-emitting ion beam for the verification of a treatment plan with a precision better than 1 mm. The maximum dose applied in the patient thereby is of the magnitude of 10 mGy.
No takes yet. Share an insight, caveat, or question.
Pawelke et al. (1996) studied this question.
Synapse has enriched 3 closely related papers on similar clinical questions. Consider them for comparative context: