Key result
Reviewing raw-data episode recordings of implantable loop recorder remote monitoring pause alerts revealed that episodes initially misdiagnosed as sinus arrest were actually high-degree AV block.
Why the study?
While implantable loop recorder remote monitoring transmissions have significant false positive alerts, true positive alerts suggesting a wrong arrhythmia diagnosis had not been shown.
Does raw-data analysis of ILR remote monitoring alerts improve the diagnostic accuracy of pause alerts compared to automated PDF reports?
Case Report (n=3)
No
Does raw-data analysis of ILR remote monitoring alerts improve the diagnostic accuracy of pause alerts compared to automated PDF reports?
Raw-data analysis of ILR remote monitoring pause alerts is essential to prevent misdiagnosing high-degree AV block as sinus arrest, which could lead to missed indications for pacemaker implantation.
Answers are grounded in this paper and cite their sources.
Raw-data review needed to distinguish high-degree AVB from sinus pauses on ILR RM alerts; hypothesis-generating for alert accuracy.
INTRODUCTION: Use of implantable loop recorder (ILR) for the diagnosis and documentation of cardiac arrhythmia has gained popularity during the last decade. Most ILRs are accompanied by remote monitoring (RM) devices, which allow automated daily interrogations of the ILR. These transmit arrhythmic or pause alerts to the treating clinic, enabling timely medical attention and treatment. Although ILR RM transmissions have been shown to have significant false positive alerts, a true positive alert suggesting a wrong arrhythmia diagnosis has not been shown to date. METHODS AND RESULTS: We revised all ILR pause transmissions arriving to our hospital device clinic during 2024, looking for pause alerts which were initially misdiagnosed. We report a case series of patients in whom a RM pause alert has been transmitted with a suspected apparent diagnosis of sinus arrest. However, when interrogating the ILR in clinic, these episodes were diagnosed as high-degree atrioventricular block (AVB) pauses, for which an urgent permanent pacemaker (PPM) might be indicated. Review of these cases revealed that in contrast with the automated RM PDF alerts, in which the figure resolution cannot be changed (and thus small P waves may be unnoticed), the raw-data episode recording on the RM web page enables one to increase the image resolution to clearly discern initially unnoticed non-conducted P waves. CONCLUSION: ILR RM pause alerts should always be reviewed via the raw-data RM web page or alternatively via ILR device interrogations, to achieve a sufficiently high-resolution image to assure these pauses are not due to AVB.
No takes yet. Share an insight, caveat, or question.
Rav et al. (2025) conducted a case report in Implantable loop recorder pause alerts (n=3). Raw-data analysis of ILR remote monitoring pause alerts vs. Standard PDF remote monitoring alerts was evaluated on Diagnosis of high-degree AV block. Reviewing raw-data episode recordings of implantable loop recorder remote monitoring pause alerts revealed that episodes initially misdiagnosed as sinus arrest were actually high-degree AV block.
Synapse has enriched 10 related papers. Consider them for comparative context: