The authors propose and demonstrate a miniaturized quadraxial probe that employs a differential feed technique for use in near-field rf/microwave transmission microscopy. Their quadraxial probe’s electric field measurements show higher electric field localization than a conventional coaxial (monopole) probe. The improved spatial resolution and more sensitive phase measurement of the quadraxial probe versus coaxial probe are further validated by a metal line scan experiment.
No takes yet. Share an insight, caveat, or question.
Karbassi et al. (2006) studied this question.