Comparative tests to measure operator protection factors in microbiological safety cabinets in accordance with British Standard 5726 have demonstrated good agreement in the results obtained by a microbiological method using a Collison nebulizer and the technique producing an aerosol of potassium iodide. Either method is suitable for testing for operator protection factors in Class I and Class II safety cabinets. The Collison nebulizer should be considered as the standard aerosol generator for the microbiological method; alternative nebulizers meeting the general requirements of BS 5726 should be compared in performance with this nebulizer if they are to be used for containment tests. Any microbiological safety cabinet specified for a new installation should have been ‘type’ tested to ensure compliance with BS 5726. However, in order to ensure adequate performance, on‐site commissioning tests (and routine maintenance checks thereafter) are necessary to verify that air velocity, filtration and operator protection factor requirements are met.