We describe an experimental method to measure the gate profile of an x-ray framing camera and to determine several important functional parameters: relative gain (between strips), relative gain droop (within each strip), gate propagation velocity, gate width, and actual inter-strip timing. Several of these parameters cannot be measured accurately by any other technique. This method is then used to document cross talk-induced gain variations and artifacts created by radiation that arrives before the framing camera is actively amplifying x-rays. Electromagnetic cross talk can cause relative gains to vary significantly as inter-strip timing is varied. This imposes a stringent requirement for gain calibration. If radiation arrives before a framing camera is triggered, it can cause an artifact that manifests as a high-intensity, spatially varying background signal. We have developed a device that can be added to the framing camera head to prevent these artifacts.