PURPOSE: To determine and track the planar orientation of active interventional devices without using localizing RF microcoils. MATERIALS AND METHODS: An image-based tracking method that determines a device's orientation using projection images was developed. An automated and a manual detection scheme were implemented. The method was demonstrated in an in vivo mesocaval puncture procedure in swine, which required accurate orientation of an active transvascular needle catheter. RESULTS: The plane of the catheter was determined using two projection images. The scan plane was adjusted automatically to follow the catheter plane, and its orientation with respect to a previously acquired target plane was displayed. The algorithm facilitated navigation for a fast and accurate puncture. CONCLUSION: Using image-based techniques, with no mechanical design changes, the orientation of an active intravascular probe could be tracked.
No takes yet. Share an insight, caveat, or question.
Sathyanarayana et al. (2007) studied this question.
Synapse has enriched 4 closely related papers on similar clinical questions. Consider them for comparative context: