This paper presents novel design, fabrication, packaging and the first in vitro neural activity recordings of SU-8-based microneedles. The polymer SU-8 was chosen because it provides excellent features for the fabrication of flexible and thin probes. A microprobe was designed in order to allow a clean insertion and to minimize the damage caused to neural tissue during in vitro applications. In addition, a tetrode is patterned at the tip of the needle to obtain fine-scale measurements of small neuronal populations within a radius of 100 µm. Impedance characterization of the electrodes has been carried out to demonstrate their viability for neural recording. Finally, probes are inserted into 400 µm thick hippocampal slices, and simultaneous action potentials with peak-to-peak amplitudes of 200–250 µV are detected.