We have developed a spatially and temporally resolved crystal spectrometer for analyzing a variety of pinch experiments on Z. The spectrometer uses a convex curved crystal to disperse spectra onto a flat microchannel plate (MCP) framing camera detector. A single wide, 1 cm, strip on the MCP is gated to provide temporal resolution. The spectral range governed by the 4 cm length of the MCP strip varies with the central Bragg angle and crystal. For a KAP crystal a typical range is 1500–2000 eV. This range can be shifted by translating the crystal along the optical axis to access different Bragg angles. The spectrometer can therefore measure K shell spectra of a wide variety of elements such as Al, Ti, and Fe. The short 1 cm width of the strip is spatially resolved with an imaging cross slit. With a 500 μm cross slit and magnification 1 the spatial resolution at the pinch is 1 mm. The instrument may also be fielded with seven time frames using a seven strip-line microchannel plate as the detector by sacrificing the spatial resolution. We will present data obtained from an aluminum pinch on Z.
No takes yet. Share an insight, caveat, or question.
Nash et al. (1999) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: