Matrix-assisted laser desorption/ionization plates coated with poly(2-hydroxyethyl methacrylate) (PHEMA) brushes that are derivatized with Fe(III)-nitrilotriacetate (NTA) complexes allow selective, efficient phosphopeptide enrichment prior to analysis by mass spectrometry (MS). Fe(III)-NTA-PHEMA brushes (60 nm thick) have a phosphopeptide binding capacity of 0.6 microg/cm(2) and exhibit phosphopeptide recoveries of over 70%, whereas much thinner polymer films containing Fe(III)-NTA afford a recovery of only 20%, and a monolayer of Fe(III)-NTA shows a recovery of just 10%. Recoveries are determined by comparing signals from enriched unlabeled phosphopeptides with those of their deuterium-labeled analogues that were added to the plate just prior to addition of matrix. Mass spectra of phosphopeptide-containing samples enriched using Fe(III)-NTA-PHEMA-modified plates also demonstrate higher recoveries or fewer interfering peaks than corresponding spectra obtained with enrichment using several commercially available Fe(III)-containing films and resins or metal oxide materials. When analyzing tryptic digests of beta-casein, the Fe(III)-NTA-PHEMA brushes allow detection of as little as 15 fmol of phosphopeptide. Moreover, with both ovalbumin and beta-casein digests, phosphopeptide signals dominate the mass spectra obtained using these modified plates.