Answers are grounded in this paper and cite their sources.
A method to measure the misalignment angles and vector fidelity of a mutually orthogonal arrangement of triaxial accelerometers has been validated by introducing known misalignments into the measurement procedure. The method is based on the excitation of all three accelerometers in equal measure and the determination of the second order responsivity tensor as a metric. The sensor axis misalignment angles measured using a sensor rotation technique as a reference were 1.49° ± 0.05°, 0.63° ± 0.02°, and 0.78° ± 0.04°. The resolution of the new approach against the reference was 0.03° with an accuracy of 0.2° and maximum deviation of 0.4°. An ellipticity tensor β that characterises the extent to which a triaxial system preserves the input polarisation state purity was introduced. In a careful laboratory arrangement, up to 98% input polarisation state purity was shown to be maintained. It is recommended that documentation on commercial and research grade high-precision triaxial sensor systems should give the responsivity matrix . This technique will improve the range of vector fidelity measurement tools for triaxial accelerometers and other vector sensors such as magnetometers, gyroscopes and acoustic vector sensors.
No takes yet. Share an insight, caveat, or question.
J M De Freitas (2018) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: