We developed a gradiometer system based on a single-layer high-temperature superconductor dc superconducting quantum interference device (SQUID), which can be freely moved within the Earth's magnetic field during measurement. The problem of circumferential shielding currents in the parallel gradiometer pick-up loop is solved by the use of an appropriately designed magnetometer SQUID integrated on the gradiometer chip. The magnetometer's feedback coil of the flux-locked loop is laid out as a small Helmholtz coil pair, thus keeping the homogeneous magnetic field constant for both the magnetometer and the gradiometer. Therefore, the balance of the directly coupled gradiometer SQUID is enhanced from 100 up to 3800. The noise limited magnetic field gradient resolution of 45 pT m −1 Hz −1/2 is preserved down to frequencies of several Hz even after strong motion in the Earth's magnetic field.