Key result
The Ension bioactive surface-treated catheter in the ETS blood loop model showed no significant elevation in blood activation markers, performing equivalently to a no-material control.
The Ension Triad System provides a high signal-to-noise model for testing blood-contacting medical devices, demonstrating that EBS surface treatment minimizes blood activation.
Supports EBS-treated catheters to limit blood activation; leaves open clinical translation from this animal model.
A novel low volume blood loop model (Ension Triad System [ETS]) incorporating pulsatile flow and a proprietary low-activation blood-contacting surface (Ension bioactive surface [EBS]) enabling high signal-to-noise performance is described. The ETS system incorporates a test chamber that allows direct comparison of material samples or finished medical devices such as catheters with varying compositions and/or surface treatments. ETS performance is presented from two independent organizations (Medtronic and MLM Labs) and includes results for hemolysis (pfHgb), platelet count, platelet activation (βTG), coagulation (TAT), inflammation (PMN Elastase, PMN CD112b, and monocyte CD112b) and immune response (SC5b-9) were made on: (1) the EBS-treated system itself without a test material (No Material, NM); (2) the EBS-treated system with an idealized untreated catheter (UC); and (3) the EBS-treated system with the prototype catheter treated with the EBS surface treatment (CC). The untreated catheter (UC) was associated with significant elevation of all activation marker levels (pfHgb excluded). The EBS-treated catheter, in direct comparison to the UC and NM catheters, appeared invisible with respect to the activation markers (all markers statistically different than the UC and equivalent to the NM control). Based on these data, we conclude that using a relatively small surface area test sample and a small volume of fresh human blood, the high signal-to-noise performance of the ETS system demonstrates comprehensive and statistically significant material differences in the major ISO 10993-4 categories of blood interaction. These data underscore the important benefit of minimal confounding of test/device responses with non-test-material/model-related responses. ETS offers a practical alternative to the common one-test-category-at-a-time approach when assessing blood/medical device interactions.
No takes yet. Share an insight, caveat, or question.
Cahalan et al. (2022) studied this question. Ension bioactive surface (EBS) treated catheter in the Ension Triad System (ETS) vs. Untreated catheter (UC) and No Material (NM) was evaluated on Blood activation markers (hemolysis, platelet count/activation, coagulation, inflammation, immune response). The Ension bioactive surface-treated catheter in the ETS blood loop model showed no significant elevation in blood activation markers, performing equivalently to a no-material control.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: