INTRODUCTION: We evaluated muscle/fat fraction (MFF) accuracy and reliability measured with an MR imaging technique at 1.5 Tesla (T) and 3.0T scanner strengths, using biopsy as reference. METHODS: MRI was performed on muscle samples from pig and rabbit species (n = 8) at 1.5T and 3.0T. A chemical shift based 2-point Dixon method was used, collecting in-phase and out-of-phase data for fat/water of muscle samples. Values were compared with MFFs calculated from histology. RESULTS: No significant difference was found between 1.5T and 3.0T (P values = 0.41-0.96), or between histology and imaging (P = 0.83) for any muscle tested. CONCLUSIONS: RESULTS suggest that a 2-point Dixon fat/water separation MRI technique may provide reliable quantification of MFFs at varying field strengths across different animal species, and consistency was established with biopsy. The results set a foundation for larger scale investigation of quantifying muscle fat in neuromuscular disorders.