A multifaceted tomotherapy quality assurance procedure has been developed. This procedure tests most of the features inherent in the tomotherapy Hi-Art device. This includes the megavoltage imaging quality, spatial and temporal accuracy of the dynamic delivery properties, as well as more traditional beam output characteristics. This is accomplished with a specialized multichannel electrometer that measures collected charge every 100 ms and a Virtual Water cylindrical phantom that holds many ion chambers and differing density insert plugs. Both devices are offered with the Hi-Art product. These tests are presented as well as their sensitivity to beam and delivery variations.
No takes yet. Share an insight, caveat, or question.
Balog et al. (2006) studied this question.