Key result
A new time-resolved fluoroimmunoassay for aCL-IgM showed a wider detectable range (0.16 to 2,630.9 MPL U/ml) and correlated well with traditional ELISA (correlation coefficient 0.955).
Effect estimate: correlation coefficient 0.955
A newly developed time-resolved fluoroimmunoassay for aCL-IgM offers a wider detectable range and high sensitivity compared to traditional ELISA for diagnosing antiphospholipid syndrome.
May expand aCL-IgM detection range in APS; leaves open prospective clinical validation.
BACKGROUND: The anticardiolipin antibodies (aCL) test has become a laboratory standard for the clinical diagnosis of antiphospholipid syndrome (APS). To better the quantitative detection of aCL-IgM so as to classify patients correctly and timely as APS positive, we established herein a new immunoassay based on a time-resolved fluoroimmunoassay (TRFIA). METHODS: The complex of cardiolipin plus bovine anti-β2 glycoprotein-I was used as antigen fixed on microtiter plates to detect serum aCL-IgM, and Eu(3+) -labeled rabbit antihuman IgM was used as conjugate. The precision, sensitivity, specificity, coefficient of recovery, and stability of the assay were evaluated, and comparison with the traditional, classical enzyme-linked immunosorbent assay (ELISA) was also made. RESULTS: The detection limit of the aCL-IgM TRFIA kit we established was 0.1 MPL U/ml, with a wider detectable range than commercial ELISA ones when a strong-positive specimen was diluted from 2,630.9 to 0.08 MPL U/ml. There was a good liner range within 0.16 to 2,630.9 MPL U/ml, whereas it was within 5.14 to 328.86 MPL U/ml when using three commercial ELISA ones. The average intra- and interassay variability was 3.19 and 3.70%, respectively. The mean recovery rate was 101.95%. The clinical diagnostic specificity was 98%. Additionally, the established assay kit presented good characteristics of stability and correlated well with the ELISA, and the correlation coefficient was 0.955. CONCLUSION: The aCL-IgM TRFIA provides an approach to a more sensitive and reliable diagnosis of APS. Further validation of its use is required.
No takes yet. Share an insight, caveat, or question.
Yan et al. (2014) studied Antiphospholipid syndrome (APS). Time-resolved fluoroimmunoassay (TRFIA) for aCL-IgM vs. Enzyme-linked immunosorbent assay (ELISA) was evaluated on Assay performance and correlation with ELISA (correlation coefficient 0.955). A new time-resolved fluoroimmunoassay for aCL-IgM showed a wider detectable range (0.16 to 2,630.9 MPL U/ml) and correlated well with traditional ELISA (correlation coefficient 0.955).
Synapse has enriched 4 closely related papers on similar clinical questions. Consider them for comparative context: