We developed a reversed-phase chromatographic procedure for detecting benzodiazepines and other drugs in serum. A liquid-liquid extraction step with hexane/ethyl acetate isolates the drugs from serum; absolute recoveries are generally greater than 85%. Reconstituted extracts are chromatographed on a 4-microns (particle size) C18 column; 14 drugs and an internal standard (flunitrazepam) are separated in 8 min. Peak detection, purity checking, and identification are performed with a computerized photodiode-array detector. Run-to-run imprecision (CV) for many benzodiazepines is less than 3%. In a study of 126 specimens from Emergency Department patients, the procedure showed excellent agreement with a gas-chromatographic method involving either mass-spectrometric or flame-ionization detection. This single procedure provides rapid and accurate detection, quantification, and confirmation of benzodiazepines in serum.
No takes yet. Share an insight, caveat, or question.
Puopolo et al. (1991) studied this question.