An optical parametric oscillator (OPO), pumped by a fiber-amplified diode laser, is combined with off-axis integrated cavity output spectroscopy (OA-ICOS). The cw OPO (power 1.2 W, tunability 3-4 μm, 5 cm(-1) mode-hop-free tuning) has a tuning speed of 100 THz/s, which is ideal for rapid and sensitive trace gas detection. Combined with OA-ICOS, a detection limit of 50 parts per trillion by volume (1×10(12)) of ethane (C(2)H(6)) in nitrogen was obtained in 0.25s at 2997 cm(-1), corresponding to a noise equivalent absorption sensitivity of 4.8×10(-11) cm(-1) Hz(-1/2). The system demonstrates real-time measurements of methane and water in exhaled human breath.
No takes yet. Share an insight, caveat, or question.
Arslanov et al. (2010) studied this question.