We have developed an optical delivery device (catheter) capable of transmitting broadband infrared light (IR wavelengths from 2 to 10 microm) for both diagnostic and therapeutic applications. The catheter is 1.68 mm in outer diameter and 1 m in length. It consists of two hollow glass waveguides coupled to a high-refractive-index optic tip. The IR light interacts with the tissue at the optic-tissue interface to measure the spectral signatures and perform therapy on the tissue at this interface. Fourier-transform IR spectrophotometer light is used to obtain the spectral signatures, and an IR free-electron laser (FEL) is used to study the therapeutic interaction of evanescent waves with the tissue. We present our catheter design; preliminary IR spectroscopy of aorta, blood, fatty tissue, and muscle; and IR FEL therapy on atheroslerotic aorta.
No takes yet. Share an insight, caveat, or question.
Hooper et al. (2003) studied this question.