Over the last 5 years, capillary zone electrophoresis (CZE) with the use of fused-silica capillaries has been increasingly introduced in clinical laboratories for routine serum-protein electrophoresis (1). The multichannel, automated Paragon CZE 2000 instrument (seven capillaries in parallel; Beckman Coulter) represented an especially attractive alternative to time-consuming manual techniques. CZE has been documented to perform reliably for the analysis of serum proteins and for the detection of monoclonal components (2)(3). We reported (4) that the sensitivity of the Paragon CZE 2000 system for the detection of monoclonal components (93%) was superior to the sensitivity of cellulose acetate gel electrophoresis (74%) and agarose gel electrophoresis (86%). In a prospective study, Katzman et al. (5) reported sensitivities of 95% and 91% for seven-capillary electrophoresis and agarose gel electrophoresis, respectively. When compared with agarose gel electrophoresis, CZE was able to detect more low-concentration IgA monoclonal components or light chains that were hidden in agarose gel electrophoresis because of comigration with transferrin or C3 (6). Paraproteins that are missed by CZE are typically very low-concentration monoclonal components that are also missed by agarose gel electrophoresis, but not by immunofixation.