Answers are grounded in this paper and cite their sources.
BACKGROUND: Schistosomiasis remains a major public health concern that afflicts millions of people worldwide. Low levels of Schistosoma infection require more sensitive diagnostic methods. In this study, a time-resolved fluoroimmunoassay (TRFIA) was developed for detecting the signal transduction protein 14-3-3, a circulating antigen of Schistosoma japonicum. RESULTS: The detection limit of 14-3-3-TRFIA was 0.78 ng/ml, with a linear measurement range from 0.78 to 800 ng/ml. The average intra-assay and inter-assay variability of this TRFIA was 8.9% and 12.2% respectively, and the mean recovery rate ranged from 92.1% to 115.5%. Within the first 21 days post-infection in rabbits, the positive rates of the 14-3-3-TRFIA were distinctly higher compared to ELISA. All these findings illustrate that 14-3-3-TRFIA has a higher detection efficacy and is a good early diagnostic method for active Schistosoma infection. CONCLUSIONS: A sandwich TRFIA for detecting the circulating antigen 14-3-3 of S. japonicum has been developed, and has demonstrated to be a good potential diagnostic method for schistosomiasis.
No takes yet. Share an insight, caveat, or question.
Qian et al. (2011) studied this question.