A fully automated instrument combining a continuous wave cavity ring-down spectrometer and dual-trap sample preconcentration has been implemented for monitoring C2H2 mixing ratios in ambient air. A distributed feedback diode laser operating in the near-infrared region (lambda approximately 1534.973 nm in air) detects C2H2 in absorption via the P(17) rotational line of the (v1 + v3) vibrational combination band. The instrument is shown to be capable of fast, quantitative, and precise monitoring of C2H2 mixing ratios, with a detection limit of approximately 8 pptv (parts per trillion by volume). It thus has potential to be deployed for analysis of air samples in many rural and urban environments. In situ measurements were carried out at 30 min intervals over periods of up to 15 h on several days for indoor and outdoor air samples. For indoor air monitored on a Sunday, the C2H2 mixing ratio was stable at 1.45 +/- 0.04 ppbv (parts per billion by volume). On weekdays, both indoor and outside air analyses showed peaks in the range 2-4 ppbv in the early morning and late afternoon that coincided with periods of busy road traffic.