We describe a procedure for measuring concentrations of cocaine in plasma by reversed-phase high-performance liquid chromatography, with ion-pairing. The procedure involves solvent extraction followed by back-extraction with dilute acid. The n-propyl ester of benzoylecgonine is used as an internal standard. An evaporation step is not required, and concentrations as low as 5 micrograms/L can be quantified. Plasma concentrations of cocaine determined by high-performance liquid chromatography correlated well with those determined by a previously reported gas-chromatographic procedure with nitrogen detection.
No takes yet. Share an insight, caveat, or question.
Jatlow et al. (1990) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: