Key result
Allegra TAVI yields ~87% technical success in high-risk valve-in-valve procedures.
Why the study?
Bioprosthetic aortic valve failure is increasing, but registry data show higher risks of coronary obstruction and elevated gradients after valve-in-valve TAVI.
Does the Allegra transcatheter heart valve system provide technical success, device success, and early safety in high-risk patients with bioprosthetic aortic valve failure undergoing valve-in-valve TAVI?
Cohort (n=106)
Yes
Does the Allegra transcatheter heart valve system provide technical success, device success, and early safety in high-risk patients with bioprosthetic aortic valve failure undergoing valve-in-valve TAVI?
The Allegra TAVI system demonstrated high technical and device success rates with excellent early safety and favorable hemodynamics for treating bioprosthetic aortic valve failure in high-risk patients.
May support Allegra feasibility in high-risk valve-in-valve TAVI; leaves open comparative efficacy pending randomized trials.
Introduction The prevalence of bioprosthetic aortic valve failure is increasing due to the broader use of bioprosthetic valves, both surgical and transcatheter, and an ageing population. Registry data indicate a higher risk of coronary obstruction and elevated gradients following valve‐in‐valve transcatheter aortic valve implantation (TAVI). We report outcomes of the Allegra transcatheter heart valve system in valve‐in‐valve TAVI. Methods In this retrospective multicentre cohort study, all patients undergoing valve‐in‐valve TAVI with the Allegra system between 2019 and 2025 at five UK centers were included. The combined primary outcomes were technical success, device success, and early safety at 30 days, as per the VARC‐3 (Valve Academic Research Consortium‐3) criteria. A key secondary outcome was all‐cause mortality at the longest documented follow‐up. Results A total of 106 patients underwent valve‐in‐valve TAVI. The mean age was 79 ± 6 years, with 57.5% being female. Almost half of the patients had an index valve ≤ 21 mm, with an STS (Society of Thoracic Surgery) score of 7.1% (5.1–12.0). The post‐implant aortic valve mean pressure gradient was 12 ± 6 mmHg, and the mean valve area was 1.7 ± 0.5 cm 2 . The combined primary outcomes were technical success of 86.8% (CI 80.3%–93.2%), device success of 75.5% (CI 67.3%–83.7%), and early safety of 86.8% (CI 80.3%–93.2%). All‐cause mortality at 30 days was 0%. All‐cause mortality occurred in 6.6% at a median follow‐up of 15 months. Conclusions The Allegra TAVI system demonstrates promising technical and device success rates, with early safety and good post‐implant haemodynamic performance in high‐risk patients undergoing valve‐in‐valve TAVI. Larger randomized trials are needed to confirm these results.
No takes yet. Share an insight, caveat, or question.
Khan et al. (2026) conducted a cohort in Bioprosthetic aortic valve failure (n=106). Allegra transcatheter heart valve system was evaluated on Technical success at 30 days (VARC-3 criteria) (95% CI 80.3%-93.2%). The Allegra TAVI system demonstrated a technical success rate of 86.8% and device success rate of 75.5% in high-risk patients undergoing valve-in-valve TAVI.