Why the study?
The incidence and impact of noise in an S-ICD after LVAD implantation was not well established.
Does LVAD implantation cause electromagnetic interference and noise in patients with a pre-existing S-ICD?
Does LVAD implantation cause electromagnetic interference and noise in patients with a pre-existing S-ICD?
LVAD implantation, particularly with HeartMate 3, causes significant electromagnetic interference in patients with pre-existing S-ICDs, often requiring the S-ICD to be permanently disabled.
Potential S-ICD noise after LVAD warrants monitoring; leaves open optimal device strategy pending larger studies.
Background: The incidence and impact of noise in a subcutaneous implantable cardioverter defibrillator (S-ICD) after left ventricular assist device (LVAD) implantation is not well established. Methods: We performed a retrospective study of patients implanted with LVAD and with a pre-existing S-ICD between January 2005 and December 2020 at the three Mayo Clinic centers (Minnesota, Arizona, and Florida). Results: Of the 908 LVAD patients, a pre-existing S-ICD was present in 9 patients (mean age 49.1 ± 13.7 years, 66.7% males), 100% with Boston Scientific third-generation EMBLEM MRI S-ICD, 11% with HeartMate II (HM II), 44% with HeartMate 3 (HM 3), and 44% with HeartWare (HW) LVAD. The incidence of noise from LVAD-related electromagnetic interference (EMI) was 33% and was only seen with HM 3 LVAD. Multiple measures attempted to resolve noise, including using alternative S-ICD sensing vector, adjusting S-ICD time zone, and increasing LVAD pump speed, were unsuccessful, necessitating S-ICD device therapies to be turned off permanently. Conclusions: The incidence of LVAD-related S-ICD noise is high in patients with concomitant LVAD and S-ICD with significant impact on device function. As conservative management failed to resolve the EMI, the S-ICDs had to be programmed off to avoid inappropriate shocks. This study highlights the importance of awareness of LVAD-SICD device interference and the need to improve S-ICD detection algorithms to eliminate noise.
No takes yet. Share an insight, caveat, or question.
Khetarpal et al. (2023) studied this question.