Answers are grounded in this paper and cite their sources.
We describe results from a multicenter pilot study evaluating the eQUANT TM system for rapid generation of a standardized inoculum (0.5 McFarland equivalent) directly from blood cultures positive for target gram-negative bloodstream pathogens for same-day antimicrobial susceptibility testing (AST) by disk diffusion, MicroScan and Vitek2 methods. Out of 167 blood cultures that met study criteria (monomicrobial with one of 9 target gram-negative species), colony counts from eQUANT suspensions of 164 (98.2 %) were within the acceptable range of 1-2e8 ± 0.6 log 10 CFU/mL (2.51×10 7 to 7.96×10 8 CFU/mL). Average eQUANT testing time was 72 min and AST results were available an average of 22.3 h earlier compared to next day testing. Overall performance of direct versus next day AST was excellent, with categorical agreement of 97.5 %, 96.9 % and 96.7 %, respectively, by disk diffusion, MicroScan and Vitek methods. For clinical application this could facilitate antibiotic stewardship and provide targeted antibiotic therapy recommendations earlier.
No takes yet. Share an insight, caveat, or question.
Abdelhamed et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: