Answers are grounded in this paper and cite their sources.
Case report demonstrates spontaneous catheter fracture and cardiac embolization after subclavian insertion, highlighting risks of cytotoxic drug extravasation.
A case of spontaneous fracture of the outlet catheter of a totally implanted catheter system (Port-A-Cath) is presented. The outlet catheter was fractured at the entrance into the left subclavian vein twenty-one weeks after insertion and the distal part was embolized in the right ventricle. The embolized catheter fragment was retrieved by a 'goose-neck' snare via the right femoral vein. The awareness of a possible spontaneous fracture of the outlet catheter of a totally implanted catheter system (Port-A-Cath) is important to prevent accidental spillage of potent cytotoxic substances.
No takes yet. Share an insight, caveat, or question.
Myint et al. (1994) studied this question.