Key result
Intraluminal pressure within a Bakri balloon during a positive tamponade test for postpartum haemorrhage does not exceed the patient's systolic blood pressure.
Why the study?
Does the intraluminal pressure within a Bakri balloon exceed systemic arterial pressure during a positive tamponade test in postpartum haemorrhage?
Case Report (n=2)
No
Does the intraluminal pressure within a Bakri balloon exceed systemic arterial pressure during a positive tamponade test in postpartum haemorrhage?
The mechanism of action of intrauterine balloons for postpartum hemorrhage may not rely on creating an intrauterine pressure greater than systemic arterial pressure.
Questions the pressure-based tamponade mechanism of intrauterine balloons in PPH; leaves open alternative actions and requires validation beyond this case report.
OBJECTIVE: To investigate the proposed mechanism by which intrauterine balloons achieve their tamponade effect of creating an 'intrauterine pressure that is greater than the systemic arterial pressure'. DESIGN: To determine the intraluminal pressures within a Bakri balloon during the establishment of a positive 'tamponade test' in the management of postpartum haemorrhage. To correlate these intraluminal pressures with contemporaneous readings of blood pressure recordings as documented from the operating theatre anaesthetic charts. SETTING: An obstetric unit (approximately 2400 births) in Wollongong, New South Wales, Australia. SAMPLE: Two women in whom first-line uterotonics were unsuccessful and who required a Bakri balloon to control postpartum haemorrhage secondary to an atonic uterus. METHODS: A DigiMano (Netech Corporation, Farmingdale, NY, USA) pressure recorder was attached via a three-way tap to a Bakri balloon. Anaesthetic charts of the two cases were reviewed retrospectively. MAIN OUTCOME MEASURES: Intraluminal pressure readings were recorded after each 50-ml aliquot of normal saline had been insufflated into the balloon whilst the next aliquot was being prepared. RESULTS: There is a curvilinear relationship between the intraluminal pressure and the balloon volume. The pressure does not exceed the systolic blood pressure of the patient at the time of establishment of a positive tamponade test. CONCLUSIONS: The intraluminal pressure within the tamponade balloon does not exceed the systolic blood pressure of the patient when a positive tamponade test is established.
No takes yet. Share an insight, caveat, or question.
Christos D. Georgiou (2009) conducted a case report in Postpartum haemorrhage (n=2). Bakri balloon was evaluated on Intraluminal pressure readings. Intraluminal pressure within a Bakri balloon during a positive tamponade test for postpartum haemorrhage does not exceed the patient's systolic blood pressure.