Key result
GyroLock, an active epicardial stabilization system using a control moment gyroscope, reduced targeted cardiac motion harmonics by 68% with accelerometers and >90% with optical sensors in vivo.
Why the study?
Does the GyroLock system reduce cardiac motion harmonics in an in vivo model of beating-heart coronary artery bypass grafting?
Does the GyroLock system reduce cardiac motion harmonics in an in vivo model of beating-heart coronary artery bypass grafting?
Answers are grounded in this paper and cite their sources.
The GyroLock system, utilizing a control moment gyroscope and adaptive control, effectively stabilizes the epicardium in vivo, offering a novel approach to facilitate beating-heart coronary artery bypass grafting.
We present herein an original solution to actively stabilize the epicardium in the context of beating-heart coronary artery bypass grafting. Our solution uses a control moment gyroscope to generate the compensation torque and an accelerometer for sensing. This approach makes the designed system completely independent from the stabilizing instrument, as well as independent from any external measurement. We compare two control approaches in a simulation: The first one uses a Kalman filter with a harmonic disturbance model, and the second one uses an adaptive algorithm. Results highlight the superiority of the adaptive control for our application. The first in vivo assessments are presented, showing the efficiency of the principle under real conditions. Using only accelerometric measurements, targeted cardiac motion harmonics are reduced on the order of 68%. With an optical sensor, the reduction exceeds 90%. This constitutes an improvement compared with prior solutions, despite the presence of nonnegligible uncertainties and distant sensing.
No takes yet. Share an insight, caveat, or question.
Gagne et al. (2012) studied Beating-heart coronary artery bypass grafting. GyroLock (control moment gyroscope) was evaluated on Reduction of targeted cardiac motion harmonics. GyroLock, an active epicardial stabilization system using a control moment gyroscope, reduced targeted cardiac motion harmonics by 68% with accelerometers and >90% with optical sensors in vivo.
Synapse has enriched 10 related papers. Consider them for comparative context: