Fast and reliable detection of endotoxin (ET) in medical equipment and pharmaceutical products is an essential precursor to clinical treatment. In this study, we demonstrate the use of shifts in wavelength of photonic crystal nanolasers for sensing the Limulus amebocyte lysate reaction, which is a standard method for detecting ET. From working curves of wavelength shift vs ET concentration, whose correlation factors were as high as 98%, we detected a required concentration of 0.001 EU/ml within 33 min and detected a low concentration of 0.0001 EU/ml.
No takes yet. Share an insight, caveat, or question.
Takahashi et al. (2015) studied this question.