Key result
Increasing spatial smoothing decreased apicoseptal strain by approximately 4% (a 13-19% proportional change), demonstrating that segmental strain values are more sensitive to smoothing settings than average four-chamber strain.
Why the study?
Clinical application of strain in neonates requires understanding which image processing parameters affect strain values, but human studies evaluating how user-regulated spatial and temporal smoothing affect neonatal strain values are lacking.
Do spatial and temporal smoothing settings affect peak systolic left ventricular longitudinal strain values in healthy neonates?
Cross-Sectional (n=36)
No
Do spatial and temporal smoothing settings affect peak systolic left ventricular longitudinal strain values in healthy neonates?
Spatial smoothing significantly alters segmental strain values in neonatal echocardiography, highlighting the need for clinicians to be mindful of software settings during assessment.
Neonatal strain values may vary with smoothing settings; leaves open optimal acquisition standards for clinical use.
Clinical application of strain in neonates requires an understanding of which image acquisition and processing parameters affect strain values. Previous studies have examined frame rate, transmitting frequency, and vendor heterogeneity. However, there is a lack of human studies on how user-regulated spatial and temporal smoothing affect strain values in 36 neonates. This study examined nine different combinations of spatial and temporal smoothing on peak systolic left ventricular longitudinal strain in 36 healthy neonates. Strain values were acquired from four-chamber echocardiographic images in the software-defined epicardial, midwall, and endocardial layers in the six standard segments and average four-chamber stain. Strain values were compared using repeated measure ANOVAs. Overall, spatial smoothing had a larger impact than temporal smoothing, and segmental strain values were more sensitive to smoothing settings than average four-chamber strain. Apicoseptal strain decreased by approximately 4% with increasing spatial smoothing, corresponding to a 13-19% proportional change (depending on wall layer). Therefore, we recommend clinicians be mindful of smoothing settings when assessing segmental strain values.
No takes yet. Share an insight, caveat, or question.
Khan et al. (2021) conducted a cross-sectional in Healthy neonates (n=36). Spatial and temporal smoothing settings vs. Low or default smoothing settings was evaluated on Peak systolic left ventricular longitudinal strain (decreased by approximately 4% (13-19% proportional change), p=<0.05). Increasing spatial smoothing decreased apicoseptal strain by approximately 4% (a 13-19% proportional change), demonstrating that segmental strain values are more sensitive to smoothing settings than average four-chamber strain.