A new tandem mass spectrometer of the quadrupole-acceleration lens-deceleration. lens-quadrupole (QADQ) configuration is described. The instrument is designed for neutralization-reionization studies and consists of a 2000-u quadrupole mass analyzer as MS-I, an acceleration electrostatic lens, a series of three differentially pumped collision cells, and an electrostatic deceleration lens, energy filter, and another 2000-u quadrupole mass analyzer as MS-II. The ion optical system achieves high total ion transmission for 5-9-keV ions. Unit mass resolution in neutralization-reionization mass spectra of aromatic compounds is demonstrated. Mass, kinetic energy, and linked scans at various levels of mass resolution and sensitivity are described.
No takes yet. Share an insight, caveat, or question.
Tureček et al. (1992) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: