A cold-cathode duoplasmatron ion source for producing beams of Ar+ or O− has been designed for the Cambridge P7-Concept microprobe secondary-ion mass spectrometer. An experimental investigation of the performance of the ion source as a function of arc current, extraction potential, magnetic field, anode to extraction-electrode spacing and, in the case of O−, intermediate electrode bore diameter and lateral offset, is presented. The conditions giving maximum source brightness have been determined. Brightness measurements give 1700 and 96 A cm−2 sr−1 for Ar+ and O−, respectively. Focused probes 4.2 μm in diameter have been produced carrying current densities of 0.1 and 0.007 A cm−2 for Ar+ and O−, respectively.
No takes yet. Share an insight, caveat, or question.
Coath et al. (1995) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: