We have constructed two types of high-temperature superconducting (HTSC) bolometers, whose performance is essentially wavelength independent from λ∼0.6 to 450 μm: The first is a microbridge, of dimensions 20×20 μm, and the second is a meander, covering a region of ∼1 mm2. Both were fabricated by photolithography of a superconducting YBaCuO thin film on SrTiO3. The bolometers are current biased, and the ac voltage induced by the chopped radiation is measured using lock-in detection. Operating optimally in the vicinity of the transition temperature (90 K), the first bolometer shows responsivity S∼1 mV/(W/cm2), while the second gives S∼800 V/W. The response of the first bolometer for various chopping frequencies is basically S∼1/√f, which makes it usable at frequencies up to ≥50 kHz, while the second bolometer has a chopping-frequency response that is basically S∼1/f, with a cutoff frequency near 0.01 Hz.
No takes yet. Share an insight, caveat, or question.
Dwir et al. (1992) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: