Answers are grounded in this paper and cite their sources.
The method used for ground-based total ozone observations and the design of two instruments used to monitor atmospheric total ozone at Arosa (Dobson spectrophotometer and Brewer spectrometer) are briefly described. Two different procedures of the calibration of the Dobson spectrometer, both based on the Langley plot method, are presented. Data quality problems that occurred in recent years in the measurements of one Dobson instrument at Arosa are discussed, and two different methods to reassess total ozone observations are compared. Two partially automated Dobson spectrophotometers and two completely automated Brewer spectrometers are currently in operation at Arosa. Careful comparison of the results of the measurements of the different instruments yields valuable information on possible small long-term drifts of the instruments involved in the operational measurements.
No takes yet. Share an insight, caveat, or question.
A 1995 study studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: