Key result
Robotic AF ablation shows zero procedural mortality but carries a notable complication rate.
Why the study?
The Hansen robotic system has only recently been used in the United States for catheter ablation procedures in humans, and early experience with its complications and technique modifications needed to be reported.
What are the complications associated with the Hansen robotic system for atrial fibrillation ablation?
Observational (n=71)
What are the complications associated with the Hansen robotic system for atrial fibrillation ablation?
Early experience with the Hansen robotic system for AF ablation highlights a learning curve with specific vascular and cardiac complications that can be mitigated through technique modifications.
Complications of early robotic AF ablation appear manageable; hypothesis-generating for safety refinements in prospective studies.
Introduction: The Hansen robotic system has only recently been used in the United States for catheter ablation procedures in humans. Atrial fibrillation (AF) ablation may be performed utilizing this system. We report our management of complications with early experience of this system. Methods and Results: All 71 patients in whom the system was utilized were included. In all patients, a 2‐operator technique was to be employed; one operator manipulates the ablation catheter via the robot and the other manipulates the circular mapping and intracardiac echocardiogram catheters. There was no procedure‐related mortality. All vascular complications occurred in the first 25 procedures performed. There were 6 intraoperative procedural‐related complications. These included significant vascular complications (n = 4), one of whom required iliac vein stenting, and 2 cardiac tamponade (one related to a pop‐phenomenon)—successfully treated by pericardiocentesis. Early complications (n = 3) were 1 tamponade several hours post‐procedure, 1 vascular complication, and 1 pericarditis. Late complications included 5 patients with severe pulmonary vein stenosis (all in first 27 patients) and 1 patient with gastroparesis. All complications were successfully managed without persistent morbidity and occurred earlier in our experience. This led to specific alterations in our vascular access and ablation techniques. These include the use of a longer 14 Fr sheath, through which the robotic sheath is more safely advanced. The choice of ablation catheter and titration of power, particularly when the catheter has a perpendicular orientation to the atrial wall, is also important. Conclusions: The suggested modifications may make the system easier to use with the potential to reduce complications.
No takes yet. Share an insight, caveat, or question.
Wazni et al. (2009) conducted an observational in Atrial fibrillation (n=71). Hansen robotic system for catheter ablation was evaluated on Procedure-related mortality and complications. In 71 patients undergoing atrial fibrillation ablation using the Hansen robotic system, there were no procedural mortalities, but 6 intraoperative, 3 early, and 6 late complications occurred.