A method for the simultaneous determination of iodine value (IV) and trans content from the Fourier transform infrared (FTIR) spectra of neat fats and oils recorded with the use of a heated single‐bounce horizontal attenuated total reflectance (SB‐HATR) sampling accessory was developed. Partial least squares (PLS) regression was employed for the development of the calibration models, and a set of nine pure triacylglycerols served as the calibration standards. Regression of the FTIR/PLS‐predicted IV and trans contents for ten partially hydrogenated oil samples against reference values obtained by gas chromatography yielded slopes close to unity and SD of <1. Good agreement (SD<0.35) also was obtained between the trans predictions from the PLS calibration model and trans determinations performed by the recently adopted AOCS FTIR/SBHART method for the determination of isolated trans isomers in fats and oils.
No takes yet. Share an insight, caveat, or question.
Sedman et al. (2000) studied this question.