Answers are grounded in this paper and cite their sources.
Well controlled cantilever oscillations are essential for precise measurements in dynamic force microscopy and spectroscopy, especially in demanding conditions such as under solution or at high frequencies. Here, we show that, with a capacitive driving system, an externally introduced DC-bias significantly increases the driving efficiency in solution and in air, as well as at high resonant frequencies, in agreement with theoretical predictions. This DC-voltage dependence not only improves the robustness of the method but also offers the possibility for precise amplitude control. The simplicity of this design further reinforces its potential for widespread implementation in dynamic force microscopy.
No takes yet. Share an insight, caveat, or question.
Zhang et al. (2013) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: