Answers are grounded in this paper and cite their sources.
This paper presents the design of an area- and energy-efficient closed-loop machine learning-based patient-specific seizure onset and termination detection algorithm, and its on-chip hardware implementation. Application- and scenario-based tradeoffs are compared and reviewed for seizure detection and suppression algorithm and system which comprises electroencephalography (EEG) data acquisition, feature extraction, classification, and stimulation. Support vector machine achieves a good tradeoff among power, area, patient specificity, latency, and classification accuracy for long-term monitoring of patients with limited training seizure patterns. Design challenges of EEG data acquisition on a multichannel wearable environment for a patch-type sensor are also discussed in detail. Dual-detector architecture incorporates two area-efficient linear support vector machine classifiers along with a weight-and-average algorithm to target high sensitivity and good specificity at once. On-chip implementation issues for a patient-specific transcranial electrical stimulation are also discussed. The system design is verified using CHB-MIT EEG database [1] with a comprehensive measurement criteria which achieves high sensitivity and specificity of 95.1% and 96.2%, respectively, with a small latency of 1 s. It also achieves seizure onset and termination detection delay of 2.98 and 3.82 s, respectively, with seizure length estimation error of 4.07 s.
No takes yet. Share an insight, caveat, or question.
Zhang et al. (2016) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: