BACKGROUND: This study was designed to evaluate the feasibility of propofol infusion by a closed-loop system for the titration of anaesthetic induction guided by Bispectral Index. METHODS: Forty patients were prospectively and randomly allocated into two groups: the target control infusion (TCI) group, where propofol titration was performed manually guided by the Bispectral Index using a commercial pharmacokinetic model (Diprifusor device) and the closed-loop group where titration was performed using a proportional differential algorithm. For both groups, the objective was to achieve a Bispectral Index of 50. Remifentanil TCI was infused at a target of 2 ng mL-1 and was maintained constant throughout the study. Feasibility of automatic induction was evaluated with performance error and haemodynamic data. RESULTS: Bispectral Index overshoot (-9 +/- 13% vs. -16 +/- 20%, P = 0.035) and mean duration of induction (381 +/- 106 s vs. 490 +/- 131 s, P = 0.004) were lower in the closed-loop group than in the TCI group. Haemodynamic data were similar between groups with a similar use of ephedrine bolus. CONCLUSION: The system was able to allow induction clinically for all patients. Automated titration guided by Bispectral Index for propofol infusion was feasible without increase in haemodynamic adverse effects.