SUMMARY As the energy of an electron beam is reduced, the range falls and the secondary electron yield rises. A low voltage scanning electron microscope can therefore, in principle, examine without damage or charging samples such as insulators, dielectrics or beam sensitive materials. This paper investigates the way in which the choice of beam energy affects the spatial resolution of a secondary electron image. It is shown that for samples which are thin compared to the electron range, the edge resolution and contrast in the image improve with increasing beam energy. In samples that are thicker than the electron range, the resolution can be optimized at either high or low energies, but low energy operation will produce images of higher contrast. At an energy of 2 keV or less beam interaction limited resolutions of the order of 3 nm should be possible.