Answers are grounded in this paper and cite their sources.
BACKGROUND: Control of the joints of robotic surgical instruments is difficult owing to hysteresis, and tendon twisting due to axial rotation of surgical tools also causes hysteresis. Therefore, a new mechanism is needed to prevent tendon twisting. METHODS: Tendon tension and hysteresis change were analysed by observing the movement of the joint depending on whether the tendons twisted for the same input signal. An anti-twist tendon mechanism to prevent twisting was developed. A 3-mm needle driver applied with the proposed mechanism was manufactured. RESULTS: The anti-twist mechanism makes no tension change because of twisting or friction between the tendon and the system, that is, the operating performance was the same regardless of rotation. CONCLUSION: The proposed mechanism has been verified and can be applied to small surgical instruments 3 mm in size. These findings can be applied in the development of instruments for precise surgeries such as microsurgery.
No takes yet. Share an insight, caveat, or question.
Kong et al. (2021) studied this question.