A novel model-based fuzzy logic controller with Kalman filtering demonstrates accurate beat-to-beat tracking of mean arterial pressure in simulation.
Simulation supports controller feasibility for MAP tracking; leaves open clinical validation in humans.
This paper proposes a new noninvasive measurement method for tracking the tendency of mean arterial pressure (MAP) in the radial artery. The designed system consists of a tonometer, a microsyringe device, and a model-based fuzzy logic controller. The modified flexible diaphragm tonometer is to detect the continuous blood pressure waveform and vessel volume pulse. A precise mathematical model describing the interaction between the tonometer and artery is derived. To reach accurate measurement without distortion, a model-based fuzzy logic control system is designed to compensate the change of MAP by applying a counter pressure on the tonometer chamber through the microsyringe device. The proposed control system consists of a linear predictor, a Kalman filter, and a synthetic fuzzy logic controller (SFLC). Simulation results show that, for the real physiologic MAP with changing rates up to 20 or -20 mm-Hg/minute, the model-based SFLC can beat-to-beat adjust the tonometer's chamber pressure to follow the tendency of MAP accurately.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2001) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: