Experimental validation reveals temperature biases and calibration needs for multi-sensor indoor air monitoring, highlighting effective correction strategies.
This paper presents the metrological characterization of low-cost sensors integrated into a multi-sensor system for Indoor Environmental Quality monitoring, developed within the MIRABLE project. The analysis focuses on two domains: the thermal domain, using Sensirion SHT45 and SEN55 temperature sensors; and the Indoor Air Quality (IAQ) domain, using an Infineon photoacoustic spectroscopy (PAS)-based sensor for carbon dioxide (CO2). All tests were conducted under controlled laboratory conditions using calibrated reference instruments. In the thermal domain, the influence of sensor integration within the device case was investigated at temperature (T) between 15 °C and 35 °C and relative humidities (RH) between 30 %rh and 60 %rh. The results revealed self-heating effects in the desk unit, causing temperature biases of up to 0.6 °C. In the IAQ domain, the repeatability and the impact of T and RH on CO2 measurements were evaluated. T was identified as the main influencing factor; whereas, RH had a negligible effect. These results were supported by statistical analysis ANOVA. A correction strategy based on concentration intervals is proposed for operation between 15 °C and 25 °C, with an expanded uncertainty (k = 2) of (3.26–6.42) ppm for the with-case configuration. These results support the reliable use of the MIRABLE system.