Key result
A tine angle of <10 degrees before the pull and hold test predicted engagement of the tines during Micra leadless pacemaker implantation (P<.0001).
Why the study?
Micra leadless pacemakers require engagement of at least 2/4 tines confirmed by a pull and hold test, but this test is sometimes difficult to assess.
Does the angle of the tines before the pull and hold test predict engagement of the tines in Micra leadless pacemaker implantation?
Cohort (n=93)
No
Does the angle of the tines before the pull and hold test predict engagement of the tines in Micra leadless pacemaker implantation?
A tine angle of <10 degrees on fluoroscopy before the pull and hold test reliably predicts tine engagement during Micra leadless pacemaker implantation.
Answers are grounded in this paper and cite their sources.
Tine angle may aid equivocal Micra deployments; hypothesis-generating and requires prospective validation before practice change.
Background: Micra leadless pacemaker is secured to the myocardium by engagement of at least 2/4 tines confirmed with pull and hold test. However, the pull and hold test is sometimes difficult to assess. This study was performed to evaluate whether the angle of the tines before the pull and hold test predicts engagement of the tines in Micra leadless pacemaker implantation. Methods: We retrospectively enrolled 93 consecutive patients (52.7% male, age 82.4 ± 9.4 years), who received Micra implantation from September 2017 to June 2020 at our institution. After deployment and before the pull and hold test, the angle of the visible tines to the body of the pacemaker was measured using the RAO view of the fluoroscopy image. The engagement of the tines was then confirmed with the pull and hold test. All tines with angles <10 degrees were engaged. In higher angles, engagement could not be predicted. Conclusion: A low angle of the tines before the pull and hold test can predict engagement of the tines in Micra leadless pacemaker implantation. The tines which are already open after deployment may be presumed that they are engaged.
No takes yet. Share an insight, caveat, or question.
Mizukami et al. (2022) conducted a cohort in Micra leadless pacemaker implantation (n=93). Angle of the tines before the pull and hold test was evaluated on Engagement of the tines confirmed with the pull and hold test (p=<.0001). A tine angle of <10 degrees before the pull and hold test predicted engagement of the tines during Micra leadless pacemaker implantation (P<.0001).