Key result
Echocardiographic particle image velocimetry is safe and feasible in patients with an LVAD, demonstrating that intraventricular left ventricular flow and vortex parameters are severely impaired.
Why the study?
Poor LV function may affect physiological intraventricular blood flow and vortex formation, prompting investigation of flow dynamics in LVAD patients via echocardiographic particle image velocimetry.
Is echocardiographic particle image velocimetry safe and feasible for quantifying intraventricular blood flow dynamics in patients with a HeartMate 3 left ventricular assist device?
Observational (n=17)
No
Is echocardiographic particle image velocimetry safe and feasible for quantifying intraventricular blood flow dynamics in patients with a HeartMate 3 left ventricular assist device?
Echocardiographic particle image velocimetry is a safe and feasible method for visualizing and quantifying the severely impaired intraventricular blood flow dynamics in patients with a HeartMate 3 LVAD.
PURPOSE: Poor left ventricular (LV) function may affect the physiological intraventricular blood flow and physiological vortex formation. The aim of this study was to investigate the pattern of intraventricular blood flow dynamics in patients with LV assist devices (LVADs) using echocardiographic particle image velocimetry. MATERIALS AND METHODS: This prospective study included 17 patients (mean age 57 ± 11 years, 82% male) who had received an LVAD (HeartMate 3, Abbott Laboratories, Chicago, Illinois, USA) because of end-stage heart failure and poor LV function. Eleven (64%) patients had ischemic cardiomyopathy, and six patients (36%) had nonischemic cardiomyopathy. All patients underwent echocardiography, including intravenous administration of an ultrasound-enhancing agent (SonoVue, Bracco, Milan, Italy). Echocardiographic particle image velocimetry was used to quantify LV blood flow dynamics, including vortex formation (Hyperflow software, Tomtec imaging systems Gmbh, Unterschleissheim, Germany). RESULTS: Contrast-enhanced ultrasound was well tolerated in all patients and was performed without adverse reactions or side effects. The LVAD function parameters did not change during or after the ultrasound examination. The LVAD flow was on average 4.3 ± 0.3 L/min, and the speed was 5247 ± 109 rotations/min. The quantification of LV intraventricular flow demonstrated substantial impairment of vortex parameters. The energy dissipation, vorticity, and kinetic energy fluctuation indices were severely impaired. CONCLUSIONS: Echo particle velocimetry is safe and feasible for the quantitative assessment of intraventricular flow in patients with an LVAD. The intraventricular LV flow and vortex parameters are severely impaired in these patients.
No takes yet. Share an insight, caveat, or question.
Schinkel et al. (2020) conducted an observational in End-stage heart failure with a left ventricular assist device (n=17). Echocardiographic particle image velocimetry vs. Previously published normal values was evaluated on Quantification of left ventricular intraventricular flow dynamics and vortex parameters. Echocardiographic particle image velocimetry is safe and feasible in patients with an LVAD, demonstrating that intraventricular left ventricular flow and vortex parameters are severely impaired.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: