An experimental study of the lubrication mechanisms existing in prototype artificial hip joints which have compliant surface linings, has shown that for a lining with a hardness of about 4N/mm2, fluid film lubrication can be achieved even with the low viscosity synovial fluid present in diseased joints after surgery. In tests, a synthetic lubricant was used in place of the synovial fluid. The best results seem to be produced with a 2 mm thick layer.
No takes yet. Share an insight, caveat, or question.
Unsworth et al. (1988) studied this question.