Answers are grounded in this paper and cite their sources.
The osseo-integration of an uncemented acetabular component depends on its initial stability. This is usually provided by under-reaming of the acetabulum. We have assessed the fixation of 52 mm porous-coated hemispherical prostheses inserted into cadaveric acetabula under-reamed by 1, 2, 3 and 4 mm. We tested the torsional stability of fixation, after preloading with 686 N in compression, by measuring the torque required to produce 1 degree and 2 degrees of rotation. Under-reaming by 2 mm and 3 mm gave significantly better fixation than 1 mm (p less than 0.01, p less than 0.02). Insertion after under-reaming of 4 mm caused some fractures. To obtain maximum interference fit and optimal implant stability, we recommend the use of an implant 2 mm or 3 mm larger than the last reamer.
No takes yet. Share an insight, caveat, or question.
Curtis et al. (1992) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: