Why the study?
Does MR angiography with superparamagnetic iron oxide (SPIO) particles provide sufficient visualization of the whole-body vascular system in adult patients?
Does MR angiography with superparamagnetic iron oxide (SPIO) particles provide sufficient visualization of the whole-body vascular system in adult patients?
MR angiography with superparamagnetic iron oxide (SPIO) particles is feasible and provides good visualization of the whole-body vascular system.
Magnetic resonance (MR) angiography with blood-pool superparamagnetic iron oxide (SPIO) particles was evaluated in the whole-body vascular system. In 12 adult patients, three-dimensional fast imaging with steady-state precession was performed in successive steps from the lungs to the calves before and after a standard dose for liver imaging (15 mumol of iron per kilogram of body weight) of AMI-25. On SPIO-enhanced MR angiograms, visualization of the pulmonary arterial, whole-body, and lower extremity venous systems was graded as good or sufficient in all patients, and femoral vein thrombosis was clearly demonstrated in one patient.
No takes yet. Share an insight, caveat, or question.
Schmitz et al. (1999) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: